chatham — Maintenance Reports
Download maintenance log
: Logger / Station Maintenance [1400]
- station maintenance, including:
- logger replacement (os downgrade: 32.07 -> 32.05)*
- logger program change (cth-bbad.cr1 -> cth-bbadC.cr1)**
* version 32.07 causes logger to reset after connecting with
LoggerNet... haven't had this issue with 32.05.
** added extra data table for debugging "phantom tips".
: Logger / Station Maintenance [1445]
- recent communication issues...
... replaced modem
- since station is on AC power...
... reset collection schedulde from 180-min to 15-min
: Logger / Station Maintenance [1500]
- routine site maintenance, including:
- replaced datalogger (to resolve faulty precipitation)
- checked / cleaned atmp/relh sensor
- checked soil sensors
- sprayed wasp nest under solar panel
- tested tipping-bucket rain gauge (good)
- weed / grass removal around station
- loaded logger program: cth-bbad.cr1* (237 -> LWS-L)*
* supports "maintenance mode"
* sensor replaced: 06/27/2022 (faulty since: 08/16/2021)
* cleaned up previously collected data
: Logger / Station Maintenance [1650]
- routine site maintenance, including:
- checked anemometer (good)
- checked atmp/relh (good)
- checked leaf wetness [standard] sensor (program update needed)*
- checked soil sensors (good)
- checked tipping-bucket rain gauge (new and level from last visit)
- checked tripod (good)
- checked voltages (good)
- checked wind vane (properly aligned)
* site does not include a canopy sensor
: Logger / Station Maintnenace: [----]
- routine site maintenance. including:
- all sensors / cables in good condition*
- weed / grass removal
* program upgrade needed to support newer LWS leaf wetness sensor
: Logger / Station Maintenance [1305]
- routine site maintenance, including:
- checked atmp/relh sensor
- checked battery (13.25 V)
- checked leaf wetness [standard] sensor (program updated needed)*
- checked pyranometer (adjusted level)
- checked soil sensor (good)
- checked solar panel (21.3 V)
- checked tipping-bucket rain gauge (adjusted level)
- cleaned atmp/relh radiation shield
- cleaned battery box
- cleaned solar panel
- removed weeds / grass under tripod
- repaired anemometer (was not connected)
- repaired wind vane (corrected alignment)
* site does not include a canopy sensor
: Logger / Station Maintenance [1000]
- routine site maintenance, including:
- checked anemometer (good)
- checked battery connections
- checked pyranometer (adjusted level)
- checked soil sensors (good)
- checked tipping-bucket rain gauge (adjusted level)
- cleaned battery box
- replaced leaf wetness (sensor) mount*
* sensor failed 08/16/2021... sensor replaced 08/26/2021
* updated logger program needed
* station does not support canopy sensor
: Logger / Station Maintenance [1305]
- routine site maintenance, including:
- replaced faulty wind vane
- replaced leaf wetness (standard) sensor [237 -> LWS]*
- replaced soil sensors (moisture and temperature)
* sensor failed 08/16/2021
* updated logger program is needed
* station will not support canopy sensor
: Logger / Station Maintenance [1505]
- replaced logger (CR1000 -> CR1000)
- added PakBus modem
- loaded logger program: cth-bbaa
: Logger Maintenance [1600]
- replaced faulty logger battery, including...
...AC power adapter
: Station Maintenance [1700]
- Loaded updatad logger program (cth-bbaa)
: Station Maintenance [1700]
- Logger upgrade from CR10X to CR1000, including:
- new modem
- new modem/lgger interface (SC-932B)
- replacement RM Young Wind Sentry
: Data Quality [1830]
- It was determined that there was a lengthy period of
faulty "solar radiation" data. All Solar Radiation
data was marked as "missing" during the following
period:
01/01/209 (d=001) 00:05 - 01/09/2013 (d=009) 11:00
: Station Maintenance [1140]
- Data cleanup performed
- Reactivated "live" station updates
: Station Maintenance [1345]
- Battery replacement... dual batteries now installed
: Station Maintenance [1204]
- Moved data-colletion to new LoggerNet Server
- Loaded Logger Program: ABCF (hourly cycle)
: Station Maintenance
- Replaced logger battery
: Station Maintenance [0002]
- Loaded Logger Program "ABCW" (3-hour collection cycle)
: Data Management [1200]
- Data cleanup performed
- Reactivated "live" updates
: Station Maintenance
- Replaced Battery. Continuing power issues have led to
missing/faulty data
: Data Management [1435]
- Due to recent troubles, data was lost for following:
12/10/2012 (d=345) 1705 - 01/09/2013 (d=009) 1100
- Reactivated "live" updates
- Noted that communication failures still occurring, although
not frequently enough to prevent adequate data collection.
: Station Maintenance [1100]
- SC932A Interface Module replaced.
- Communication re-established.
: Station Maintenance
- Logger replaced
: Station Maintenance [0005]
- Loaded MAWN Program - ABCF (hourly collection cycle)
: Station Maintenance [0002]
- Loaded MAWN Program - ABCW (three-hour collection cycle)
: Station Maintenance [0006]
- Loaded MAWN Program - ABCF (hourly collection cycle)
: Station Maintenance [0006]
- Loaded MAWN Program - ABCW (3-hour collection cycle)
: Station Maintenance [1345]
- Recent logger/communication issues have led to a loss of
data. Unfortunately, data was not reported/collected
during the following time period:
03/24/10 (d=083) 15:40 - 04/07/10 (d=097) 15:20
: Station Maintenance
- Loaded MAWN Program - ABCF (hourly collection cycle) [0002]
: Station Maintenance
- Loaded MAWN Program - ABCW (3-hour collection cycle) [0001]
: Station Maintenance [1200]
- On 08/11/09 at 14:00, communication to/from the Chatham
datalogger was severed. Rather than install/bury new
short-haul modem cable, this opportunity was used to
upgrade the Chatham station to the CELL-IP Network. A
wireless broadband modem was installed before noon and
data-collection/monitoring is now controlled from the
EnviroWeather Office on the campus of Michigan State
University.
- MAWN Program - ABCF Loaded (60-minute power-cycle).
- Added Hourly Battery Voltage to MAWN Database tables
for Chatham.
: Station Maintenance
- Snowfall adapter was removed.
: Station Maintenance
- Recent telecommunications with the data-collection PC
for the Chatham station have been repaired. As of
10:10am this morning, data is flowing once again. No
data was lost during the temporary outage.
: Station Maintenance
- Snowfall adapter for Tipping Bucket Rain Gauge Installed.
NOTE: Timing and amount of precip over the past couple of days
may be slightly off as there was some frozen precipitation
prior to the installation of the snowfall adapter. It
was reported that 1.44" of precipitation fell between
10/22/06 - 10/26/06.
: Station Maintenance
- Snowfall adapter for Tipping Bucket Rain Gauge Installed.
: Station Maintenance (sm)
- General Station Maintenance and System Checks:
- repaired leaf wetness sensor
- cleaned Precipitation bucket
- checked Air Temp/Relative Humidty sensor
: Station Maintenance (sm)
- General Station Maintenance and System Checks:
- Checked Temperature/Relative Humidity Sensor (HMP45C). No need
to replace as this sensor was calibrated in 03/2004.
- Checked Tipping Bucket Rain Gauge [10:45].
- Repaired Communication Wiring. Wires used to connect short-haul
modem to the back of the PC were not connected properly. Repaired
and reconnected.
- Work was done approx: 10:30 - 11:00 (Day=275)
: Data Collection and Data Archive for new station begins
First report: Jun 18, 2003 - 00:05 (Day=169)
: New Station Installed
Notes:
- leaf wetness B (canopy) - still needs to be installed.
- Data Logger Program Loaded: MAWN1C
Additional Station Needs:
- Pyronometer stand needs leveling plate. Sensor is
installed and tied down, but without proper leveling
plate.