- replaced wind sentry (wind speed / wind direction)

- replaced wind sentry (wind speed / wind direction)
- repaired tipping-bucket rain gauge (reinstalled)
- power system upgrade
- replaced / remounted leaf wetness (standard mount)
- several issues... replaced the following:
- logger and solid state relay
- atmp/relh sensor (HMP155A)
- all soil sensors
- solar radiation (pyranometer)
- tipping-bucket rain gauge
- all leaf wetness sensors
- replaced logger (CR1000 -> CR1000 / PakBus change)
- loaded logger program: epr-bbcd
- replaced faulty pyranometer (solar radiation sensor)
- loaded logger program (epr-bbcd):
- corrected issue with missing canopy leaf wetness
- corrected issue with "air temp" sensor change
- site rebuild...
- replaced logger (CR1000)
- replaced atmp/relh (HMP45C -> HMP155A)
- replaced leaf wetness (237 -> LWS-L)
- replaced soil temp sensors
- replaced soil moisture sensors
- loaded logger program (epr-bbab)*
* needs to be upgraded to "epr-bbcd"
- replaced faulty cellular modem
- replaced faulty logger battery
- replaced logger battery
- replaced logger battery
- replaced logger battery
- replaced faulty CR10X logger with CR1000 logger
- loaded logger program "mawn_abcf" (hourly collection cycle)
- faulty logger battery replaced
- moved data-collection to new LoggerNet server
- loaded logger program "mawn_abcw"
- routine maintenance
- Loaded Logger Program "ABCF" (hourly collection cycle)
- Loaded Logger Program "ABCW" (3-hour collection cycle)
- Replaced logger battery.
- Battery issues led to recent data loss.
- Replaced Battery. Battery issues led to recent data loss
- Replaced Battery. Battery issues led to recent data loss
- Removed blockage from rain gauge funnel.
- Will need to mark all precipitation data as "missing"
during the following time period:
09/04/2012 (d=248) 00:00 - 09/17/2012 (d=261) 18:10
- Replaced Modem
- Replaced Battery
- Battery issues led to data loss during following periods:
12/15/2011 (d=349) 1705 - 12/17/2011 (d=351) 1000
12/17/2011 (d=351) 1705 - 12/18/2011 (d=352) 1300
12/18/2011 (d=352) 1705 - 12/20/2011 (d=354) 1820
- Loaded MAWN Program - ABCF (hourly collection cycle)
- Loaded MAWN Program - ABCW (3-hour collection cycle)
- Loaded MAWN Program - ABCF (hourly collection cycle) [0004]
- Loaded MAWN Program - ABCW (3-hour collection cycle) [0005]
- Recent communication troubles...
- Wireless broadband modem replaced on 06/07/09.
- LoggerNet software adjusted on 06/08/09.
- Telecommunication trouble with the Eastport MAWN
station was corrected - blown fuse in modem power
cable was replaced.
- Removed faulty data introduced due to battery failure.
There is now a large chunk of several missing data
entries between 11/18/08 - 12/08/08.
- Replaced faulty battery [1726]
- Loaded MAWN Program - ABCF (60-minute power cycle) [0006]
- Tipping Bucket rain gauge was repaired (cleaned) [1430]
- Communication converted to broadband cell-modem:
- Broadband cell-ip modem installed
- Datalogger progam "mawn_abcs" loaded and compiled.
- Battery voltage now reported every hour.
- Data collection configured to upload every 30 minutes.
- General station maintenance and system checks.
- Work was done approx: 14:35 (Day=134)
- Phone line to datalogger was accidentally cut on Friday,
May 27, 2005 and was repaired today - phone line was
spliced at gated entrance to the orchard and buried for
*temporary* protection.
- Data collection and data upload to MAWN database resumed
normal operation.
- General Station Maintenance and System Checks:
- weed removal and application of Round-Up above soil temps
- cleaned Air Temp/Relative Humidty sensor
- check Anemometer/Wind Vane
- cleaned Precipitation bucket
- New Datalogger Program (MAWN1C) Loaded (16:30, Day=103).
The new program corrects soil moisture readings which
were previously reported as missing.
First report: April 05, 2004 - 00:05 (Day=096)
Notes:
- leaf wetness B (canopy) - sensor is installed, but stii
needs to be installed in tree canopy.
Additional Station Needs: